2.1.4. UPMODE Command
The UPMODE
command allows to select or read the setpoint Update Mode.
Command Format:
R/W |
Command |
Response |
---|---|---|
R |
UPMODE:? |
#UPMODE:<mode> |
W |
UPMODE:<mode> |
#AK / #NAK |
Parameter(s):
<mode> |
Description |
---|---|
NORMAL |
Setpoint is set using standard Ethernet communication |
WAVE |
Setpoint is taken from the internal function generator module |
AIN |
Setpoint is set using analog input [-10V,+10V] (available only when optional board are installed - see Optional Boards) |
SFP |
Setpoint is taken from a dedicated custom command sent to SFP interface |
Tip
When AIN mode is selected, the relative Analog input signal [-10V,+10V] is converted to the module full-scale value. So for example when a module has a full-scale of 150A - 10V, then in CC mode the +10V is is converted to +150A; otherwise in CV mode the +10V remains +10V.
Example(s):
UPMODE:WAVE
#AK
UPMODE:?
#UPMODE:WAVE