5.2. Fault register
The Fault Register indicates the unit’s fault register. To read it, see the REG Command.
Bit(s) |
Name |
Fault Type |
Description |
---|---|---|---|
[0] |
Temperature Fault |
Soft |
Indicates a over-temperature fault or that the temperature sensor is disconnected |
[1] |
DC-Link Fault |
Hard |
DC-Link fault |
[2] |
Input OVC |
Hard |
Input Over-Current fault |
[3] |
Over-Power |
Soft |
Over-power fault |
[4] |
Earth-Leakage |
Soft |
Earth-Leakage fault - see Internal Memory id:115 |
[5] |
Earth Fuse |
Soft |
Earth Fuse fault |
[6] |
Regulation Fault |
Soft |
Regulation fault - see Internal Memory id:120-122 |
[7] |
Reserved |
/ |
/ |
[8] |
DCCT Error |
Hard |
DCCT Error |
[9] |
Reserved |
/ |
/ |
[10] |
Crowbar |
Hard |
Crowbar Fault |
[11] |
Quench |
Soft |
Quench Detection Fault - see Internal Memory id:130-134 |
[12] |
Reserved |
/ |
/ |
[13] |
Output OVC |
Hard |
Output Over-Current Fault - Output current exceeded OVC Threshold (see Internal Memory id: 100) |
[14] |
Output OVV |
Hard |
Output Over-Voltage Fault - Output voltage exceeded OVV Threshold (see Internal Memory id: 101) |
[15] |
Reserved |
/ |
/ |
[16] |
External Interlock #1 |
Soft |
External Interlock #1 Fault (see INTERLOCK Command) |
[17] |
External Interlock #2 |
Soft |
External Interlock #2 Fault (see INTERLOCK Command) |
[18] |
External Interlock #3 |
Soft |
External Interlock #3 Fault (see INTERLOCK Command) |
[19] |
External Interlock #4 |
Soft |
External Interlock #4 Fault (see INTERLOCK Command) |
[23:20] |
Reserved |
/ |
/ |
[24] |
Parallel Config. Error |
Soft |
Model or Firmware mismatch between the Master and Slave unit(s) |
[25] |
Parallel Link Fault |
Hard |
Communication Error between the Master and Slave unit(s) |
[26] |
Reserved |
/ |
/ |
[27] |
Fan(s) Fault |
Soft |
Fan(s) Fault |
[31:28] |
Reserved |
/ |
/ |
See also
For a detailed description of the faults, please refer to the User’s Manual