List of the Error Codes
The CT-BOX unit replies with a Not AcKnowledge (NAK) if the received
command is not correct, or is not accepted. The NAK reply is followed by an
error code field, which indicates the error cause and/or type.
Error Code |
Brief description of error |
|---|---|
|
Command not valid |
|
Parameter not valid |
|
Parameter is out of expected range |
|
Parameter cannot be modified when the CT-BOX is acquiring data |
|
Printout stream has been interrupted (DLOG/DMM only): repeat the issued command or restart the printout stream with PRINT:ON |
|
Command is not supported in the current mode (see MODE) |
|
Command cannot be executed when the CT-BOX is acquiring data |
|
RS-232 communication is only supported in DLOG mode |
|
DCCT is not connected to the CT-BOX |
|
Other DCCT error |
|
CT-BOX name parameter string is too long (see CTBOX) |
|
CT-BOX is already acquiring data |
|
SD-Card file create error |
|
SD-Card file close error |
|
SD-Card write to file error |
|
SD-Card file cannot be synchronized |
|
SD-Card is not mounted (see SD) |
|
SD-Card cannot list files |
|
SD-Card cannot be mounted |
|
SD-Card cannot be unmounted |
|
It is not possible to remove the file |
|
It is not possible to open the file |
|
It is not possible to read the file |
|
Password required (see PASSWORD) |
|
|
|
Wrong password |
|
USB communication is only supported in DLOG mode |
|
Trigger mode is not valid |