PTURNS Command
Applies to: ALL modes (general purpose).
The PTURNS (Primary Turns) command sets the number of primary turns. When
the primary conductor is wound N times through the hole of the DCCT,
the DCCT measures N times the conductor current. This increases the
sensitivity to small currents, while the full-scale current of the conductor
is divided by N.
The full-scale current depends on the connected DCCT (see DCCT). As an example, with a 200 A DCCT the full-scale current of the conductor is 100 A with two primary turns, and about 66.7 A with three primary turns.
Warning
Do not apply the rated nominal full-scale primary current (for example 600 A for a CT-600) when carrying out multiple turns on the primary conductor hole.
The value N given to the command has to be equal to the number of turns
physically applied to the DCCT, so that the CT-BOX divides the
readings by the same factor and reports the current flowing in the conductor.
The factory-defined number of turns is 1 (no turns applied to the primary
conductor).
Note
The number of primary turns is stored in the unit’s flash memory and is
retained after a power cycle. Set it back to 1 when the additional turns are
removed from the DCCT.
Command Format:
R/W |
Command |
Response |
Description |
|---|---|---|---|
W |
|
|
Set the number of turns applied to the DCCT |
R |
|
|
Read the configured number of turns |
Parameter(s):
Name |
Type |
Allowed values |
Default |
Description |
|---|---|---|---|---|
|
int |
1 – 100 |
|
Number of primary turns |
Example(s):
Set the number of turns to 3:
PTURNS:3
ACK
PTURNS:?
3