GETT Command
Applies to: ALL modes (general purpose).
The purpose of the GETT command is to read the temperature of the
DCCT or of the connected external temperature sensor.
The format of the returned temperature is always expressed as an ASCII
floating point number with one decimal digit, in degrees
Celsius [°C], terminated by \r\n.
Command Format:
R/W |
Command |
Response |
Description |
|---|---|---|---|
R |
|
|
Read the temperature of the DCCT |
R |
|
|
Read the temperature of the external temperature sensor |
Note
If the DCCT is not connected, GETT:HEAD returns -9999.0.
Likewise, if the external sensor is not connected, GETT:EXT returns
-9999.0.
Note
The external temperature sensor is attached to the D-SUB DA-15S (female) connector on the back of the CT-BOX unit; refer to the CT-BOX User’s Manual for the connector pinout.
Example(s):
GETT:HEAD
36.7
GETT:EXT
25.3