Command Table Summary
The following table lists every CT-BOX command, its purpose, its parameter(s) and the mode(s) of operation in which it applies, for quick reference.
In the Mode column: ALL means a general-purpose command usable in all three
acquisition modes, OSC means the command applies only in oscilloscope mode, DMM means the command applies only in digital-multimeter mode,
DLOG means it applies only in data-logger mode, DLOG-DMM means it applies in both data-logger and digital-multimeter modes, DLOG-OSC means it applies in both data-logger and oscilloscope modes.
Settings of mode-specific commands can be written and read in any mode; only
GET and ACQT are rejected outside their modes. Data is saved to the SD card only in DLOG
and DMM modes.
Mode and Acquisition
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Set the device in OSC / DLOG / DMM mode |
|
ALL |
Read the device operation mode |
|
ALL |
|
|
Start / stop the measurements acquisition |
|
ALL |
Read the acquisition status |
|
ALL |
|
|
Enable / disable printing on the communication stream |
|
ALL |
Read the printing configuration |
|
ALL |
Reading
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Get a single current reading, in Amps |
(none) |
DLOG-OSC |
|
Get an array of |
|
ALL |
|
Read the DCCT / external sensor temperature |
|
ALL |
Output Data Rate and Acquisition Settings
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Set the averaging interval, in µs |
|
OSC |
Read the averaging interval |
|
OSC |
|
|
Set the output data frequency, in Hz |
|
DLOG |
Read the output data frequency |
|
DLOG |
|
|
Set the DMM submode |
|
DMM |
Read the DMM submode |
|
DMM |
|
Set the DMM frequency |
|
DMM |
|
Read the DMM frequency |
|
DMM |
|
Set the DMM number of periods |
|
DMM |
|
Read the DMM number of periods |
|
DMM |
|
|
Set the temperature acquisition configuration |
|
DLOG |
Read the temperature acquisition configuration |
|
DLOG |
Measurement Setup
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Set the number of primary turns |
|
ALL |
Read the configured number of primary turns |
|
ALL |
|
|
Zero the offset at null primary current |
|
ALL |
Alarm and Trigger
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Enable / disable the alarm check functionality |
|
ALL |
Clear the alarm condition |
|
ALL |
|
Read the alarm check setting |
|
ALL |
|
Set / read the upper current threshold, in Amps |
|
ALL |
|
Set / read the lower current threshold, in Amps |
|
ALL |
|
|
Set the trigger as input / output / gate or disable it |
|
IN / OUT: DLOG-OSC; GATE: DMM; OFF: ALL |
Read the trigger status |
|
ALL |
SD Card
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Mount / unmount the SD card |
|
DLOG-DMM |
Read the SD card status |
|
DLOG-DMM |
|
Reset the SD card |
|
DLOG-DMM |
|
List all files on the SD card |
|
DLOG-DMM |
|
Remove / size / read a file on the SD card |
|
DLOG-DMM |
|
|
Enable / disable saving to the SD card |
|
DLOG-DMM |
Read the saving configuration |
|
DLOG-DMM |
Real-Time Clock
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Set / read the RTC date ( |
|
DLOG-DMM |
|
Set / read the RTC time ( |
|
DLOG-DMM |
Network
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Set / read the IP address |
|
ALL |
|
Set / read the subnet mask |
|
ALL |
|
Set / read the default gateway |
|
ALL |
|
Set / read the UDP data-stream port |
|
OSC |
Device Information and Reset
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Read the installed firmware release |
|
ALL |
|
Read the unit serial number |
|
ALL |
Set / read the user-defined unit name (max 15 chars) |
|
ALL |
|
Read the unit calibration date |
|
ALL |
|
|
Read the matched DCCT serial number |
|
ALL |
Read the connected DCCT model and serial number |
|
ALL |
|
|
Perform a hardware reset of the unit |
|
ALL |
Status and Errors
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Read the internal Status register |
|
ALL |
|
Read the internal Error register |
|
ALL |
Clear the internal Error register |
|
ALL |
Calibration (Password Protected)
Command |
Purpose |
Parameters |
Mode |
|---|---|---|---|
|
Unlock / lock the protected commands |
|
ALL |
|
Calculate the calibration parameters |
|
ALL |
|
Store / read the calibration parameters |
|
ALL |